Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21680 |
| Brand Name: | AxiEM Surgical Procedure Kit, General Purpose, Reusable |
| Model: | |
| Affected: | Straight Suction 9733449 EM ENT, Curved Suction 70 9733450 EM ENT, Curved Suction 90 9733451 EM ENT, Suction 9734308 Small Straight AxiEM ENT |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Navigation Inc |
| Issue: | The manufacturer has become aware that the specific lots of AxiEMTM ENT Suction instruments are not able to pass instrument verification. Instrument verification is a step in the software which is performed prior to the use of each instrument. This verification step confirms device tracking functionality and prevents use of the device if a passing verification is not achieved. The impacted devices were made with a particular stainless steel stock material that exhibits magnetic characteristics (magnetic permeability), causing interference with electromagnetic (EM) tracking capability of the StealthStationTM. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 12/7/2017 |