Detail of Recall

Type of Product:Device
Medsafe Reference:21680
Brand Name:AxiEM Surgical Procedure Kit, General Purpose, Reusable
Model:
Affected:Straight Suction 9733449 EM ENT, Curved Suction 70 9733450 EM ENT, Curved Suction 90 9733451 EM ENT, Suction 9734308 Small Straight AxiEM ENT
Software version:
Recalling Organisation:Medtronic New Zealand Limited
Unit 16 Mezzanine Level
5 Gloucester Park Road
Onehunga
AUCKLAND
Contact Information:0800 377 807
Manufacturer:Medtronic Navigation Inc
Issue:The manufacturer has become aware that the specific lots of AxiEMTM ENT Suction instruments are not able to pass instrument verification. Instrument verification is a step in the software which is performed prior to the use of each instrument. This verification step confirms device tracking functionality and prevents use of the device if a passing verification is not achieved. The impacted devices were made with a particular stainless steel stock material that exhibits magnetic characteristics (magnetic permeability), causing interference with electromagnetic (EM) tracking capability of the StealthStationTM.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:12/7/2017