Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:21676
Brand Name:Alere HIV Combo
Model:
Affected:20 Tests Kits catalogue number 7D2846, Lot number 80671K100C
Software version:
Recalling Organisation:Abbott Laboratories (NZ) Ltd - Rapid Diagnostics Divison
Ground Floor, Bldg. D
4 Pacific Rise
Mount Wellington
AUCKLAND 1060
Contact Information:0800 466 951
Manufacturer:Alere Medical Co Ltd
Issue:Internal investigations have identified that when testing with EDTA whole blood samples, may exhibit increased frequency of a red bar in the antigen (p24) result window, which could be interpreted as a false positive result. These bars may be faint. Antigen (p24) performance with other sample types continues to perform as per product specifications. Performance for HIV-1 and HIV-2 antibody is as per product specifications.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:11/7/2017