Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:21666
Brand Name:Philips DigitalDiagnost DXR
Model:
Affected:DigitalDiagnost model 712220
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems DMC GmbH
Issue:A firmware anomaly inside the detector (4343RC) can cause the detector to appear ready for acquisition on very short time intervals while it actually is not, resulting in not correctly acquiring the X-ray image.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/7/2017
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /