Detail of Recall

Type of Product:Device
Medsafe Reference:21663
Brand Name:XN-L series analyser
Model:
Affected:XN-L series instruments: version 00-13 and above
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Sysmex Corporation
Issue:It has been reported that the WBC could become falsely low when the following steps are performed:
Manual mode is selected
Whole blood mode (WB), PreDilute mode (PD) or body fluid mode (BF) is changed to low WBC mode (LW).
The measurement information is input in the manual order dialogue and the menu is closed by selecting 'Ok'.
No measurement is performed in LW mode, and the mode is then changed to use another measurement mode (ie either WB, PD, BF).
With the next measurement, the WBC result becomes diluted and gives a result half of the correct WBC value.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/7/2017