Detail of Recall

Type of Product:Device
Medsafe Reference:21660
Brand Name:Dimension/Dimension Vista Systems
Model:
Affected:Alanine Aminotransferase, Aspartate Aminotransferase, Glucose, Creatine Kinase MB and Thyroxine assays All Lots
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
55 Hugo Johnston Drive
Penrose
Auckland
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has become aware of sulfasalazine and sulfapyridine drug interference in the assays which use NADH and/or NADPH to generate reduction oxidation reactions which produce colorimetric signals. Falsely depressed or falsely elevated results may occur on samples drawn from patients taking Sulfasalazine and Sulfapyridine
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/7/2017