Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:21658
Brand Name:ADVIA Chemistry Systems
Model:
Affected:ADVIA Chemistry Products affected by Sulfasalazine and Sulfapyridine, ADVIA Chemistry Products affected by Sulfasalazine only
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has become aware of sulfasalazine and sulfapyridine drug interference in the assays which use NADH and/or NADPH to generate reduction oxidation reactions which produce colorimetric signals. It has been confirmed that falsely depressed or falsely elevated results may occur on samples drawn from patients taking Sulfasalazine and Sulfapyridine
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/7/2017