Detail of Recall

Type of Product:Device
Medsafe Reference:21610
Brand Name:AllSet = Gold SSP
Model:
Affected:Lots 54340D (Lot 038), 54030D (Lot 005), 54330D (Lot 021)
Software version:
Recalling Organisation:Bio-Strategy Ltd
22a William Pickering Drive
North Harbour
Albany
Auckland 0751
Contact Information:09 969 9150
Manufacturer:Thermo Fisher Scientific Australia Pty Ltd
Issue:The manufacturer is aware of the AllSet+ Gold SSP kits. The primers PM074D and C15-02A should be negative for the following alleles: :, B*07:219, B*57:02:01, , B*57:02:02, , B*57:03:02, , B*57:07, B*57:09, B*57:12, B*57:17, B*57:28N, B*57:39, B*57:42, B*57:46, B*57:57, B*57:63, B*57:66, B*57:70, B*57:80, B*57:82, B*58:08:01, B*58:08:02, B*58:20, B*58:27, B*58:28:01, B*58:28:02, B*58:65, and B*58:73 during analysis. The Lot specific user documents and software UCH files contain the incorrect allele specificity information for these two primer mixes.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:3/7/2017