Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:21595
Brand Name:Sensis / Sensis Lite
Model:
Affected:Utilizing Windows XP and Windows 7 operating systems
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Brand Name:Artis Q, Q.zen, zeego, zee systems
Model:
Affected:Utilizing Windows XP and Windows 7 operating systems
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Brand Name:Artis one
Model:
Affected:Utilizing Windows XP and Windows 7 operating systems
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Shenzhen Magnetic Resonance Ltd
Brand Name:ARCADIS Avantic/Orbic/Varic, Cios Connect & Siremobile Compact
Model:
Affected:Utilizing Windows XP and Windows 7 operating systems
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare GmbH
Issue:The WannaCry ransomware attack was a worldwide cyberattack by the WannaCry ransomware cryptoworm, which targeted computers running the Microsoft Windows operating system by encrypting data and demanding ransom payments in the Bitcoin cryptocurrency. Many of the systems are utilizing Windows XP and Windows 7 operating systems and therefore are at risk from WannaCry virus. The manufacturer has released security patches as correction to address this issue, however the medical device is still performing according to its specifications and security patches are only cybersecurity enhancements.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/6/2017
Hide menus
Show menus