Detail of Recall

Type of Product:Device
Medsafe Reference:21594
Brand Name:Balloon pump
Model:
Affected:CS100 and CS300 Intraortic Balloon Pumps.
Software version:
Recalling Organisation:Maquet Australia Pty Ltd (NZ)
12G Andromeda Cres East Tamaki
Auckland 2013
Contact Information:021 874 882
Manufacturer:Datascope Corporation
Issue:The manufacturer has received a complaint involving a CS300 IABP that did not pump due to an electrical test failure code #58 (power up vent tests fail), maintenance code #3, and an autofill failure. An electrical test failure code #58 is caused by a solenoid valve requiring more power than the solenoid driver board can deliver to open the valve. The lack of power prevents the coil from moving the plunger causing the valve not to open.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/6/2017