Detail of Recall

Type of Product:Device
Medsafe Reference:21593
Brand Name: Warsaw - Sterile Inst
Model:
Affected:BALL TIP GUIDE WIRE 2.4MM, 3MM X 70CM TEAR DROP GUIDE WIRE
Software version:
Recalling Organisation:Zimmer Biomet New Zealand Company
210 Khyber Pass Road
Grafton
Auckland
Contact Information:09 925 5200
Manufacturer:Zimmer Inc
Issue:The design verification for the previous packaging configuration does not cover the 70cm wires. While there has not been any complaints involving a breach in sterility, the manufacturer has completed design verification to move the 70cm guide wires to a new packaging configuration. As a result, the products packaged in the previous packaging configuration are being removed
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/6/2017