Detail of Recall

Type of Product:Device
Medsafe Reference:21590
Brand Name:Cobas modular analytic series system
Model:
Affected:Immunochemistry module cobas e 602, cobas e 601 and E170 modules, cobas 8000 modular analyzer series, cobas 6000 analyzer series, MODULAR ANALYTICS, MODULAR ANALYTICS EVO
Software version:
Recalling Organisation:Roche Consumer Health, Division of Roche Products (NZ) Limited
8 Henderson Place
Te Papapa
AUCKLAND
Contact Information:
Manufacturer:Patheon API Inc.
Issue:In very rare cases when a number of specific conditions are met, it is possible that a sample mismatch may occur in the immunochemistry modules (cobas e 602, cobas e 601 and E170 modules) due to a software limitation. If the Gripper fails to pick up a sample vessel (cup) for a sample on the rack being assayed, the measurement for that sample is cancelled and the warning "Tip/Cup pick up error" (Caution Level) is issued.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:28/6/2017