Detail of Recall

Type of Product:Device
Medsafe Reference:21587
Brand Name:ACS900 800 Flowtron TriPuls
Model:
Affected:Software V1.099, Serial Numbers range from 1400028031 to 1600048470
Software version:
Recalling Organisation:Arjo New Zealand Limited
34 Vestey Drive
Mt Wellington
AUCKLAND 1060
Contact Information:0800 000 151
Manufacturer:ArjoHuntleigh AB
Issue:It has been identified that in some instances the pump will constantly deliver pressure to the garment without triggering the alarm. The manufacturer has investigated the symptom encountered and has concluded that the consistent flow of pressure to the garment without deflation may lead to a potential risk of injury to patients.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:27/6/2017