Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:21582
Brand Name:Sysmex XN 10 and Xn 20
Model:
Affected:XN-10: all instruments, XN-20: all instruments
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Sysmex Corporation
Issue:The manufacturer has informed that misalignment of the protective sheet on Sysmex XN-10 and XN-20 analysers may lead to salt deposits, which could potentially build-up on the electronic components. In rare circumstances, this salt deposit build-up over time may become an internal combustion hazard.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/6/2017
Hide menus
Show menus
0 1 2 4 5 6 7 9 [ /