Type of Product:Device
Medsafe Reference:21580
Brand Name:cobas e411 immunoassay analyser
Model:
Affected:All models (Disk and Rack), all software versions.
Software version:
Recalling Organisation:Roche Diagnostics (NZ) Ltd
15 Rakino Way
Mt Wellington
AUCKLAND 1644
Contact Information:0800 363 622
Manufacturer:Patheon API Inc.
Issue:In very rare cases a software malfunction in the Sample&Control datafile can occur which may lead to a potential data mismatch. This software malfunction only occurs when the "Sample Data Clear" function is not performed daily as indicated in the Operator's Manual, and when the Sample&Control datafile is filled with more than 2000 records.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:23/6/2017