Detail of Recall

Type of Product:Device
Medsafe Reference:21574
Brand Name:LifePoint bPLUS
Model:
Affected:Models: LifePoint Pro, LifePoint bPLUS, LifePoint mPLUS
Software version:
Recalling Organisation:Total Safety Solutions Ltd
19 Bridgehead Cove
Hobsonville
Auckland 0618
Contact Information:
Manufacturer:Metsis Medikal Ltd Sti
Issue:As a result of the momentary communication interruption caused by the software of Automatic External Defibrillator devices during the self-test procedure, the device got stuck in stand-by during this process and the BIOS battery was depleted then the device was locked-up due to processor being subjected to low supply voltage during depletion of BiOS battery and the warning indicator was stuck in GREEN, not switching to RED by full depletion of BIOS battery. The self-test is performed by the main processor and cannot control the indicator position in the event of possible instantaneous communication and lockup problems.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/6/2017