Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21574 |
| Brand Name: | LifePoint bPLUS |
| Model: | |
| Affected: | Models: LifePoint Pro, LifePoint bPLUS, LifePoint mPLUS |
| Software version: | |
| Recalling Organisation: | Total Safety Solutions Ltd 19 Bridgehead Cove Hobsonville Auckland 0618 |
| Contact Information: | |
| Manufacturer: | Metsis Medikal Ltd Sti |
| Issue: | As a result of the momentary communication interruption caused by the software of Automatic External Defibrillator devices during the self-test procedure, the device got stuck in stand-by during this process and the BIOS battery was depleted then the device was locked-up due to processor being subjected to low supply voltage during depletion of BiOS battery and the warning indicator was stuck in GREEN, not switching to RED by full depletion of BIOS battery. The self-test is performed by the main processor and cannot control the indicator position in the event of possible instantaneous communication and lockup problems. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/6/2017 |