Detail of Recall

Type of Product:Device
Medsafe Reference:21572
Brand Name:Sysmex CA-600 series system
Model:
Affected:Sysmex CA-620 and Sysmex CA-660
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Sysmex Corporation
Issue:Under rare circumstances there is a potential that the system might fail to alert the operator about an individual Sample ID No. assigned to multiple patient measurement results when using the built-in barcode scanner. Under very rare circumstances this may lead to an incorrect diagnostic and patient treatment decision due to potential of a mismatched result for the patient.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/6/2017