Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21568 |
| Brand Name: | CS - Adaptor, luer (DLP) |
| Model: | |
| Affected: | DLP Extension Line Adapters 20 in length, DLP Antegrade/Retrograde Adapter, DLP Cardioplegia Adapter with Pressure Port, Multiple Lots |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Brand Name: | CS - Tubing set, pressure monitoring, invasive blood pressure |
| Model: | |
| Affected: | DLP Pressure Monitoring Extension Line Adapters, DLP Pressure Disposable Pressure Display Sets , Multiple Lots |
| Software version: | |
| Recalling Organisation: | Medtronic New Zealand Limited Unit 16 Mezzanine Level 5 Gloucester Park Road Onehunga AUCKLAND |
| Contact Information: | 0800 377 807 |
| Manufacturer: | Medtronic Inc |
| Issue: | During the manufacturing process, the manufacturer identified small pinholes, approximately 1 mm or less in diameter, in a single packaging configuration of sterile pouches. Based on internal investigation, all products in this packaging configuration manufactured since March 1, 2016 are potentially affected. The result is a potential sterile barrier breach. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 21/6/2017 |