Type of Product:Device
Medsafe Reference:21564
Brand Name:IGUIDE 2.2.0
Model:
Affected:using iGUIDE 2.2.0, iGUIDE 2.2.1
Software version:
Recalling Organisation:Elekta Pty Limited - New Zealand Branch
Suite 302, Geyser Building, 100 Parnell Road
Parnell
AUCKLAND 1052
Contact Information:0273-544-793
Manufacturer:Elekta Inc
Issue:If the PEC data dialog after a verification scan is cancelled, a re-send of PEC data will lead to incorrect positioning because the 2nd PEC values supersede the 1st PEC values instead of being added.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:22/6/2017