Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21564 |
| Brand Name: | IGUIDE 2.2.0 |
| Model: | |
| Affected: | using iGUIDE 2.2.0, iGUIDE 2.2.1 |
| Software version: | |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Elekta Inc |
| Issue: | If the PEC data dialog after a verification scan is cancelled, a re-send of PEC data will lead to incorrect positioning because the 2nd PEC values supersede the 1st PEC values instead of being added. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 22/6/2017 |