Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21549 |
| Brand Name: | Radspeed Manual / Auto |
| Model: | |
| Affected: | Tube Support, CH-200/CH-200M product range |
| Software version: | |
| Recalling Organisation: | Shimadzu Medical Systems NZ Ltd 9-5 Northside Drive Whenuapai Auckland 0814 |
| Contact Information: | |
| Manufacturer: | Shimadzu Corporation |
| Issue: | It was found that a crack may occur over time on the tube holding shaft near the base of the tube mounting flange. If these cracks occur and then increase in size, the tube mounting flange may separate from the tube holding shaft. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 16/6/2017 |