Type of Product:Device
Medsafe Reference:21549
Brand Name:Radspeed Manual / Auto
Model:
Affected:Tube Support, CH-200/CH-200M product range
Software version:
Recalling Organisation:Shimadzu Medical Systems NZ Ltd
9-5 Northside Drive
Whenuapai
Auckland 0814
Contact Information:
Manufacturer:Shimadzu Corporation
Issue:It was found that a crack may occur over time on the tube holding shaft near the base of the tube mounting flange. If these cracks occur and then increase in size, the tube mounting flange may separate from the tube holding shaft.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/6/2017