Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21547 |
| Brand Name: | Portable ventilator, electric |
| Model: | |
| Affected: | All Astral 100 and Astal 150 ventilators |
| Software version: | |
| Recalling Organisation: | ResMed (NZ) Ltd 266 Runciman Road Pukekohe Auckland |
| Contact Information: | 0800 737 635 |
| Manufacturer: | ResMed (Australia) Ltd |
| Brand Name: | Power supply, battery pack |
| Model: | |
| Affected: | All Astral External Batteries and All Astral battery packs (spares) |
| Software version: | |
| Recalling Organisation: | ResMed (NZ) Ltd 266 Runciman Road Pukekohe Auckland |
| Contact Information: | 0800 737 635 |
| Manufacturer: | ResMed (Australia) Ltd |
| Issue: | The manufacturer has received a small number of report regarding the performance of the internal battery, where the false activation of a fuse leads to shut down of the internal battery. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Healthcare Professional |
| Date Commenced: | 19/6/2017 |





