Type of Product:Device
Medsafe Reference:21516
Brand Name:Oscor Electrode Cable
Model:
Affected:Reuseable ATAR multiple lots
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Oscor Incorporated
Issue:In some cases during the use of some ATAR™ reusable extension cables, the cables was separating from the connector at the proximal end. The analysis of the returned devices revealed that the connector and wire separation was attributed to a fracture of the conductor cable most likely caused by extensive use of the reusable cables beyond the recommended maximum amount of 3 times. This event resulted in cable malfunction, causing interruption of the pacing system.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/6/2017