Detail of Recall

Type of Product:Device
Medsafe Reference:21491
Brand Name:Infinity Acute Care System Monitoring - M540
Model:
Affected: Kontron C500 and C700 Cockpits
Software version:
Recalling Organisation:Draeger New Zealand Ltd
Unit 4
24 Bishop Dunn Place
East Tamaki
Auckland 2013
Contact Information:0800 559 186
Manufacturer:Draegerwerk AG & Co KGaA
Issue:The Virtium RAM may induce a failure that can lead to patient risk (safety issue). It has been determined that the Swissbit 4 GB RAM Modules perform as intended. The Virtium RAM may induce a failure that can lead to high memory usage thereby delaying touchscreen operations, and also may lead to not annunciating audio or visual alarms on the Cockpit only.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/6/2017