| Type of Product: | Device |
| Medsafe Reference: | 21491 |
| Brand Name: | Infinity Acute Care System Monitoring - M540 |
| Model: | |
| Affected: | Kontron C500 and C700 Cockpits |
| Software version: | |
| Recalling Organisation: | Draeger New Zealand Ltd Unit 4 24 Bishop Dunn Place East Tamaki Auckland 2013 |
| Contact Information: | 0800 559 186 |
| Manufacturer: | Draegerwerk AG & Co KGaA |
| Issue: | The Virtium RAM may induce a failure that can lead to patient risk (safety issue). It has been determined that the Swissbit 4 GB RAM Modules perform as intended. The Virtium RAM may induce a failure that can lead to high memory usage thereby delaying touchscreen operations, and also may lead to not annunciating audio or visual alarms on the Cockpit only. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 7/6/2017 |