Detail of Recall

Type of Product:Device
Medsafe Reference:21489
Brand Name:Hospira - Plum 360 Pump
Model:
Affected:List Number 30010
Software version:
Recalling Organisation:ICU Medical Australia Pty Ltd (NZ)
Level 7
The Bayleys Building
36 Brandon Street
Wellington 6011
Contact Information:
Manufacturer:Hospira Inc
Issue:The manufacturer has identified a potential for the Connectivity Engine (CE) Module to disengage from the main chassis. In a rare situation, this could lead the Plum 360 Infuser to power down without an alarm notification resulting in a delay of either initiation of a therapy or interruption of an active infusion.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:6/6/2017