Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21489 |
| Brand Name: | Hospira - Plum 360 Pump |
| Model: | |
| Affected: | List Number 30010 |
| Software version: | |
| Recalling Organisation: | ICU Medical Australia Pty Ltd (NZ) Level 7 The Bayleys Building 36 Brandon Street Wellington 6011 |
| Contact Information: | |
| Manufacturer: | Hospira Inc |
| Issue: | The manufacturer has identified a potential for the Connectivity Engine (CE) Module to disengage from the main chassis. In a rare situation, this could lead the Plum 360 Infuser to power down without an alarm notification resulting in a delay of either initiation of a therapy or interruption of an active infusion. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 6/6/2017 |