Detail of Recall

Type of Product:Device
Medsafe Reference:21481
Brand Name:Surgical Tissue Pouches
Model:
Affected: Lapsac Introducer, G16500 All Lots, G15771 All Lots
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Cook Incorporated
Issue:The reprocessing instructions do not provide sufficient detailed information for the cleaning, disinfection, and sterilisation of these products. Preliminary investigation indicates that validation data related to the reprocessing of these devices do not meet the current guidance.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:1/6/2017