Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:21476
Brand Name: DxH 500
Model:
Affected:Software Versions 1.0.2
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has become aware that when the instrument temperature (<18°C to >36.5°C) is exceeded, the DxH 500 allows sample processing and result reporting despite the Instructions for Use stating that The run sample is inaccessible. Results generated in this scenario may be erroneous. Additionally:
" The message on the results screen/printout reads Instrument Temperature instead of Instrument Temperature Out of Range.
" The results are transmitted to the Laboratory Information System (LIS) without the accompanying Instrument Temperature Out of Range message. Using the auto-transmit option could lead to the release of erroneous results to the LIS.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/5/2017