Detail of Recall

Type of Product:Device
Medsafe Reference:21475
Brand Name:ABL90Flex analyser
Model:
Affected:Cord blood results
Software version:
Recalling Organisation:Radiometer Pacific
Suite 1
1/30 Benedicts Street
Eden Terrace
Auckland 1010
Contact Information:0800 723 722
Manufacturer:Radiometer Medical ApS
Issue:The ABL90 series analyzers can report on either cord blood arterial or cord blood venous. However, irrespectively of the sample type selected on the ABL90 series analyzer during sample processing, the analyzer will transmit the less specific sample type of cord blood to a HIS/LIS or middleware system.
Recall action type:Correction
Recall action:Software to be upgraded
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:7/6/2017