Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21473 |
| Brand Name: | Carestream - DRX Revolution |
| Model: | |
| Affected: | Serial numbers 101 to 2623 |
| Software version: | |
| Recalling Organisation: | Carestream Health New Zealand Ltd c/- Quigg Partners Level 7, 28 Brandon St Wellington |
| Contact Information: | 021 977 215 |
| Manufacturer: | Carestream Health Inc (HCIS) |
| Issue: | The manufacturer has received a report of the tube and yoke assembly of a DRX Revolution Mobile X-Ray System separating. This is caused by 4 missing mounting screws in the assembly – not installed at the time of manufacturing. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 31/5/2017 |