Type of Product:Device
Medsafe Reference:21473
Brand Name:Carestream - DRX Revolution
Model:
Affected:Serial numbers 101 to 2623
Software version:
Recalling Organisation:Carestream Health New Zealand Ltd
c/- Quigg Partners
Level 7, 28 Brandon St
Wellington
Contact Information:021 977 215
Manufacturer:Carestream Health Inc (HCIS)
Issue:The manufacturer has received a report of the tube and yoke assembly of a DRX Revolution Mobile X-Ray System separating.
This is caused by 4 missing mounting screws in the assembly – not installed at the time of manufacturing.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:31/5/2017