Detail of Recall

Type of Product:Device
Medsafe Reference:21471
Brand Name:Calprotectin Card (20 test kit with stool vials & positive control)
Model:
Affected:Lots CP-052
Software version:
Recalling Organisation:Abacus ALS Limited
55b Richard Pearse Drive
Airport Oaks
Auckland 2022
Contact Information:0800 222 170
Manufacturer:Abacus Diagnostics
Issue:Lots CP-052 and CP-047 were evaluated using a strip reader and standard preparations of Calprotectin ranging from 125 ng/mL to 4000 ng/mL. Comparative results using the strip reader and visual examination showed the intensity of the reactions were stronger on lot CP-047 than lot CP-052. When the concentration of calprotectin in a stool sample is near to the test detection limit (500 ng/mL), some samples could be interpreted as negative, depending on the lot and the visual interpretation.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:30/5/2017