Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21467 |
| Brand Name: | UniCel DxI System |
| Model: | |
| Affected: | Access Total T3, Access Thyroglobulin, Access Free T4, Access GI Monitor All Lots |
| Software version: | |
| Recalling Organisation: | Beckman Coulter NZ Ltd Suite 2, Level 4 30 St Benedicts Street Newton AUCKLAND |
| Contact Information: | 0800 446 109 |
| Manufacturer: | Beckman Coulter Inc |
| Issue: | The manufacturer has determined through customer feedback and an internal investigation that the four Access immunoassays are susceptible to biotin interference. During interference testing the interference occurred with samples that contained 100 ng/mL of biotin. This level of biotin is greater than the maximum biotin concentration observed in the normal healthy population. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 24/5/2017 |