Detail of Recall

Type of Product:Device
Medsafe Reference:21467
Brand Name:UniCel DxI System
Model:
Affected:Access Total T3, Access Thyroglobulin, Access Free T4, Access GI Monitor All Lots
Software version:
Recalling Organisation:Beckman Coulter NZ Ltd
Suite 2, Level 4
30 St Benedicts Street
Newton
AUCKLAND
Contact Information:0800 446 109
Manufacturer:Beckman Coulter Inc
Issue:The manufacturer has determined through customer feedback and an internal investigation that the four Access immunoassays are susceptible to biotin interference. During interference testing the interference occurred with samples that contained 100 ng/mL of biotin. This level of biotin is greater than the maximum biotin concentration observed in the normal healthy population.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/5/2017