| Type of Product: | Device |
| Medsafe Reference: | 21434 |
| Brand Name: | Philips V60 Ventilator |
| Model: | |
| Affected: | Internal cable for all V60 Ventilators with a date of manufacture before 15 September 2015 |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Respironics California LLC |
| Issue: | Over time, low-frequency vibrations can cause the pins within the female connectors on the internal Motor Controller to Data Acquisition Board ribbon cable to become partially displaced, which causes momentary high resistance that interferes with data transfer. This may cause the ventilator to fail Power on Self Testing (POST) or cause Continuous Built in Test (CBIT) to detect a fault and lead to a ventilator shut down with alarm during use or during intra-hospital transport. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/5/2017 |