Detail of Recall

Type of Product:Device
Medsafe Reference:21434
Brand Name:Philips V60 Ventilator
Model:
Affected:Internal cable for all V60 Ventilators with a date of manufacture before 15 September 2015
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Respironics California LLC
Issue:Over time, low-frequency vibrations can cause the pins within the female connectors on the internal Motor Controller to Data Acquisition Board ribbon cable to become partially displaced, which causes momentary high resistance that interferes with data transfer. This may cause the ventilator to fail Power on Self Testing (POST) or cause Continuous Built in Test (CBIT) to detect a fault and lead to a ventilator shut down with alarm during use or during intra-hospital transport.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/5/2017