Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21428 |
| Brand Name: | Anaesthesia Work Stations |
| Model: | |
| Affected: | Carestation 620, 650 and 650c |
| Software version: | |
| Recalling Organisation: | GE Healthcare Limited Level 7, Vero Centre 48 Shortland St Auckland 1010 |
| Contact Information: | 0800 659 465 |
| Manufacturer: | Datex-Ohmeda Inc |
| Issue: | An incomplete seal can exist between the disposable absorber and the breathing circuit lower assembly of the Carestation 600 Series systems. This incomplete seal can allow rebreathing of patient gases that have bypassed the Carbon Dioxide (CO2) Absorbent material and could result in unintended elevated inspired levels of CO2 (FiCO2), which could lead to hypercarbia. |
| Recall action type: | Correction |
| Recall action: | Instructions for use to be updated |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 18/5/2017 |