Detail of Recall

Type of Product:Device
Medsafe Reference:21428
Brand Name:Anaesthesia Work Stations
Model:
Affected: Carestation 620, 650 and 650c
Software version:
Recalling Organisation:GE Healthcare Limited
Level 7, Vero Centre
48 Shortland St
Auckland 1010
Contact Information:0800 659 465
Manufacturer:Datex-Ohmeda Inc
Issue:An incomplete seal can exist between the disposable absorber and the breathing circuit lower assembly of the Carestation 600 Series systems. This incomplete seal can allow rebreathing of patient gases that have bypassed the Carbon Dioxide (CO2) Absorbent material and could result in unintended elevated inspired levels of CO2 (FiCO2), which could lead to hypercarbia.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:18/5/2017