Type of Product:Device
Medsafe Reference:21426
Brand Name:Asserachrom VWF: CB
Model:
Affected:Lot 114710(RE)
Software version:
Recalling Organisation:
Contact Information:
Manufacturer:
Issue:The Asserachrom® VWF:CB lot 114710(RE) gives higher results compared to the other Asserachrom ® VWF:CB tested lots for samples with a VWF:CB level above 30%. This issue is more pronounced for samples with a VWF:CB level above 50%.
Based on our internal testing,
Normal samples tested are overestimated by 26% on average (relative deviation). For example: a normal plasma at 134% could be reported at 167%.
The pathological samples tested and in particular the type 1 plasmas are overestimated by 38% on average (relative deviation).
For example: a type 1 plasma at 56% could be reported at 75%.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:16/5/2017