Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21403 |
| Brand Name: | Cook Inc (Urology) Holmium Laser Fibres (Multi Use) (HLF-M) |
| Model: | |
| Affected: | Catalogue number G48611 HLF-M273-CN, G48613 HLF-M365-CNV, G48615 HLF-M550-CNV All Lots |
| Software version: | |
| Recalling Organisation: | Obex Medical Ltd Level 3, 109 Carlton Gore Road Newmarket Auckland 1023 |
| Contact Information: | 0800 656 539 |
| Manufacturer: | Cook Incorporated |
| Issue: | The manfacturer has identified that the reprocessing instructions do not provide sufficiently detailed information for the cleaning, disinfection, and sterilization of these products. The manufacturer preliminary investigation indicates that the validation data related to the reprocessing of these devices do not meet the current guidance. |
| Recall action type: | Recall |
| Recall action: | Product to be returned to supplier |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 5/5/2017 |