Detail of Recall

Type of Product:Device
Medsafe Reference:21403
Brand Name:Cook Inc (Urology) Holmium Laser Fibres (Multi Use) (HLF-M)
Model:
Affected: Catalogue number G48611 HLF-M273-CN, G48613 HLF-M365-CNV, G48615 HLF-M550-CNV All Lots
Software version:
Recalling Organisation:Obex Medical Ltd
Level 3, 109 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:0800 656 539
Manufacturer:Cook Incorporated
Issue:The manfacturer has identified that the reprocessing instructions do not provide sufficiently detailed information for the cleaning, disinfection, and sterilization of these products. The manufacturer preliminary investigation indicates that the validation data related to the reprocessing of these devices do not meet the current guidance.
Recall action type:Recall
Recall action:Product to be returned to supplier
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:5/5/2017