Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:21394
Brand Name:ARCHITECT Free T4 Reagent
Model:
Affected:List number 7K65-29, 100 Test Kit Multiple lots, 7K65-34, 2000 Test Kit Multiple lots, 7K65-39, 500 Test Kits Muliple lots
Software version:
Recalling Organisation:Abbott Australasia Pty Ltd (NZ)
Abbott Laboratories NZ
4 Pacific Rise, Mt Wellington
Auckland 1060
Contact Information:61 2 93849717
Manufacturer:Abbott Bioresearch Centre
Issue:The manufacturer has determined that the measuring interval is 0.40 ng/dL to 5.00 ng/dL. the estimated frequency of falsely elevated or decreased fT4 results is occasional (0.04%). Only results well above the medical decision point are impacted. Changes to results in this range of values are unlikely to change clinical interpretation and therefore, patient management is not impacted.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Wholesale
Date Commenced:1/5/2017