Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21393 |
| Brand Name: | IBA Stealth Chamber DS50-0030XL-T |
| Model: | |
| Affected: | IBA DS50-003-XL-T Stealth Chamber |
| Software version: | |
| Recalling Organisation: | Alphatech Systems Limted 630a Great South Road Greenlane Auckland |
| Contact Information: | 09 580 1959 |
| Manufacturer: | IBA Dosimetry GmbH |
| Issue: | When the Stealth Chamber is used as a reference detector in relative dosimetry, deviation in the measured PDDs at shallow depths and geometric shadowing of measured profiles may appear. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 1/5/2017 |