Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21356 |
| Brand Name: | Medical Intelligence HexaPOD evo RT System |
| Model: | |
| Affected: | Hexapod IGUIDE 2.1, IGUIDE 2.2 |
| Software version: | |
| Recalling Organisation: | Elekta Pty Limited - New Zealand Branch Suite 302, Geyser Building, 100 Parnell Road Parnell AUCKLAND 1052 |
| Contact Information: | 0273-544-793 |
| Manufacturer: | Medical Intelligence Medizintechnik GmbH |
| Issue: | If by mistake the initial pre-treatment imaging was performed BEFORE the HexaPOD was moved to the DRIVE (START) position, iGUIDE offers the possibility to branch into a specific catch-up workflow, which was implemented to use the initial scan data. This should avoid the need to redo imaging with HexaPOD in DRIVE (START) position. In this scenario the software may prompt you several times to move the HexaPOD to its DRIVE (START) position. |
| Recall action type: | Correction |
| Recall action: | Software to be upgraded |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 28/4/2017 |