Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21355 |
| Brand Name: | Infusion pumps - Syringe (Swiss) |
| Model: | |
| Affected: | Product codes with prefix (all variants): 8001, 8002, 8003, 8004, 8005 |
| Software version: | |
| Recalling Organisation: | CareFusion New Zealand 313 Limited Unit 4 / 110 Mays Road Onehunga AUCKLAND 1061 |
| Contact Information: | 0508 422 734 |
| Manufacturer: | CareFusion Switzerland 317 Sarl |
| Issue: | The manufacturer has identified that a breakage of the plunger backplate spring may allow movement of the syringe plunger within the plunger holder mechanism; this may result in siphonage. In some circumstances, this may result in a clinically significant over infusion. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 27/4/2017 |