Database updated: 27 October 2026

Safety Information

Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:21346
Brand Name:Bard Australia Pty Limited Arctic Sun
Model:
Affected:ARCTIC SUN 5000 Temperature Management System
Software version:
Recalling Organisation:Pharmacy Retailing (NZ) Ltd t/a Healthcare Logistics
54 Carbine Road
Mt Wellington
AUCKLAND
Contact Information:09 969 0705
Manufacturer:Medivance Inc
Issue:The ARCTIC SUN® 5000 Temperature Management System when in excess of 2,000 system hours, without conducting preventative maintenance, may result in failure of certain components in the system, specifically issues associated with insufficient cooling and pumping issues.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:26/4/2017
Hide menus