Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21346 |
| Brand Name: | Bard Australia Pty Limited Arctic Sun |
| Model: | |
| Affected: | ARCTIC SUN 5000 Temperature Management System |
| Software version: | |
| Recalling Organisation: | Pharmacy Retailing (NZ) Ltd t/a Healthcare Logistics 54 Carbine Road Mt Wellington AUCKLAND |
| Contact Information: | 09 969 0705 |
| Manufacturer: | Medivance Inc |
| Issue: | The ARCTIC SUN® 5000 Temperature Management System when in excess of 2,000 system hours, without conducting preventative maintenance, may result in failure of certain components in the system, specifically issues associated with insufficient cooling and pumping issues. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 26/4/2017 |