Brand Name:Siemens Dimension V-LYTE® Dilution Check and V-LYTE® Integrated Multisensor Technology (IMT)
Model:
Affected:V-LYTE Dilution Check, V-LYTE Integrated Multisensor Technology (IMT)
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has identified two issues related to the V-LYTE® Integrated Multisensor Technology (IMT) System:
a) Dilution Check Relative Bias Failures - a high rate of dilution check relative bias failures of greater than 2% when using V-LYTE® Dilution Check solution.
b) Integrated Multisensor (IMT) Urine Sodium Drift - elevated urine sodium Quality Control (QC) exceeding acceptable QC ranges.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:24/4/2017