Detail of Recall

Type of Product:Device
Medsafe Reference:21340
Brand Name:SD BIOLINE HIV Ag/Ab Combo, SD03FK30
Model:
Affected:Lot numbers, 03BDA005A, 03BDA006A, 03BDA006B, 03BDB001C
Software version:
Recalling Organisation:Abbott Laboratories (NZ) Ltd - Rapid Diagnostics Divison
Ground Floor, Bldg. D
4 Pacific Rise
Mount Wellington
AUCKLAND 1060
Contact Information:0800 466 951
Manufacturer:Alere Medical Co Ltd
Issue:The manufacturer received complaints and confirmed that certain lots did not meet the performance claim for analytical sensitivity of p24 antigen (the claimed detection limit in the package insert is 2 IU/mL, however the lots listed detect a concentration of 16 IU/mL.)
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:20/4/2017