Detail of Recall

Type of Product:Device
Medsafe Reference:21303
Brand Name:Dihydrotestosterone (DHT) RIA
Model:
Affected:Cat. Number. DE5170, Lot. 164808/C
Software version:
Recalling Organisation:Abacus ALS Limited
55b Richard Pearse Drive
Airport Oaks
Auckland 2022
Contact Information:0800 222 170
Manufacturer:
Issue:The manufacturer noticed that the KMnO4 reagent (#3205213) - lot 16H8 - included in the Dihydrotestosterone (DHT) RIA kit may lead to elevated concentrations, especially for male samples. The impact of the affected KmnO4 (lot 16H8) was less visible on female samples. Because this reagent is used in the extraction step, elevated values were not observed on the Standard Curve nor on the Kit Controls. Only extracted samples are susceptible to higher results.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/4/2017