Medsafe Online Recall Database

Detail of Recall

Type of Product:Device
Medsafe Reference:21295
Brand Name:Arrow International - Peripherally-inserted central venous catheter
Model:
Affected:ARROW PICC Kits with Peel-Away Sheath Component
Software version:
Recalling Organisation:Teleflex Medical New Zealand
OFFICE 011, 110 CARLTON GORE ROAD
Newmarket
Auckland 1023
Contact Information:027 476 4009
Manufacturer:Arrow International Inc
Issue:The manufacturer has received complaints for peel away sheaths flaring. The instructions for use included in the kits do not inform the user that a skin nick or alternate enlargement method may be required for difficult insertions. The use of a skin nick reduces the probability of sheath flare and subsequent complications.
Recall action type:Correction
Recall action:Instructions for use to be updated
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:10/4/2017