Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21265 |
| Brand Name: | Philips Angiographic X-Ray |
| Model: | |
| Affected: | Allura Xper and Allura CV20 Releases 3 to 8.1 with an installed Laird chiller, Allura Xper Release 8.2: only configurations FD10, FD10/10, FD20/10, FD20/20 have a Laird chiller |
| Software version: | |
| Recalling Organisation: | Philips New Zealand Commercial Limited Level 3 123 Carlton Gore Road Newmarket AUCKLAND 1023 |
| Contact Information: | 0800 251 400 |
| Manufacturer: | Philips Medical Systems Nederland BV |
| Issue: | Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R-cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odour, smoke or fire. |
| Recall action type: | Correction |
| Recall action: | Product to be modified |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 29/3/2017 |