Detail of Recall

Type of Product:Device
Medsafe Reference:21265
Brand Name:Philips Angiographic X-Ray
Model:
Affected:Allura Xper and Allura CV20 Releases 3 to 8.1 with an installed Laird chiller, Allura Xper Release 8.2: only configurations FD10, FD10/10, FD20/10, FD20/20 have a Laird chiller
Software version:
Recalling Organisation:Philips New Zealand Commercial Limited
Level 3
123 Carlton Gore Road
Newmarket
AUCKLAND 1023
Contact Information:0800 251 400
Manufacturer:Philips Medical Systems Nederland BV
Issue:Due to a leak in the detector cooling system, cooling liquid may leak outside the drip tray of the chiller. If this occurs the liquid may drip onto electrical components in the R-cabinet located in the technical room, which could lead to damage to the system and potentially cause thermal events such as a burning odour, smoke or fire.
Recall action type:Correction
Recall action:Product to be modified
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:29/3/2017