Database updated: 27 October 2026
Safety Information
Medsafe Online Recall Database
Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21238 |
| Brand Name: | Biomet UK - Uni Knee |
| Model: | |
| Affected: | OXF UNI C/LESS TIB TRAYS LM size AA through to OXF UNI C/LESS TIB TRAY RM size F |
| Software version: | |
| Recalling Organisation: | Biomet New Zealand Ltd |
| Contact Information: | 0275 522 595 |
| Manufacturer: | Biomet UK Ltd |
| Issue: | The Oxford Cementless Tibial Trays have been manufactured according to pre-defined specifications. An investigation into the cause of the reported tibial plateau fractures has revealed the importance that certain steps described in the applicable surgical technique are followed to reduce the risk of tibia plateau fractures. |
| Recall action type: | Alert |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Healthcare Professional |
| Date Commenced: | 20/3/2017 |