Detail of Recall

Type of Product:Device
Medsafe Reference:21235
Brand Name:ADVIA Centaur BR assay (50 tests) and (250 test)
Model:
Affected: Catalogue numbers – 10333349 and 10334837
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed the potential for the ADVIA Centaur Syphilis assay to interfere with the BR (CA 27.29) assay on the ADVIA Centaur Systems. This interference only occurs when the Syphilis assay immediately precedes the BR assay on the ADVIA Centaur, ADVIA Centaur XP or ADVIA Centaur XPT systems. The ADVIA Centaur CP system is not impacted.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Wholesale
Date Commenced:17/3/2017