Detail of Recall

Type of Product:Device
Medsafe Reference:21213
Brand Name:Roche Cobas c
Model:
Affected:Cobas c311: Catalogue No. 05899427001, cobas c501, c502: Catalogue No. 04547241001, cobas c702, cobas 8000 ISE: Catalogue No. 04945794001
Software version:
Recalling Organisation:Roche Products (NZ) Ltd
98 Carlton Gore Road
Newmarket
Auckland 1023
Contact Information:
Manufacturer:Roche Diagnostics GmbH
Issue:In very rare cases where the disturbance of the sample liquid level detection (LLD) occurs, the affected sample probe may dip into the sample material deeper than intended, the affected sample probe may be not washed adequately. This may lead to carryover.
Recall action type:Recall
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:14/3/2017