Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21203 |
| Brand Name: | ETEST Ceftriaxone TXL32 FOAM packaging |
| Model: | |
| Affected: | ETEST Ceftriaxone TXL32 (Ref. 507018) |
| Software version: | |
| Recalling Organisation: | Medi'Ray New Zealand Ltd 53-55 Paul Matthews Road Albany Auckland 0632 |
| Contact Information: | 0800 633 472 |
| Manufacturer: | Biomerieux SA |
| Issue: | The manufacturer has observed that the current shelf-life claims of the ETEST Ceftriaxone TXL32 products are not supported by internal testing. As a result there is a potential for an overestimation of the MIC values for specifically Neisseria gonorrhoeae. |
| Recall action type: | Recall |
| Recall action: | Product to be destroyed |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 10/3/2017 |