Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21202 |
| Brand Name: | Visumax |
| Model: | |
| Affected: | battery circuit for the UPS |
| Software version: | |
| Recalling Organisation: | Carl Zeiss (NZ) Ltd 22 Fremlin Place Avondale Auckland 1026 |
| Contact Information: | 09 838 5626 |
| Manufacturer: | Carl Zeiss Meditec AG |
| Issue: | Complaints in the past have indicated that in certain devices there is an increased probability of failure of the battery test circuit components of the UPS-module. Therefore the battery circuit for the UPS has been modified to make the UPS more robust. |
| Recall action type: | Correction |
| Recall action: | Product to be exchanged |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 9/3/2017 |