Detail of Recall

Type of Product:Device
Medsafe Reference:21202
Brand Name:Visumax
Model:
Affected:battery circuit for the UPS
Software version:
Recalling Organisation:Carl Zeiss (NZ) Ltd
22 Fremlin Place
Avondale
Auckland 1026
Contact Information:09 838 5626
Manufacturer:Carl Zeiss Meditec AG
Issue:Complaints in the past have indicated that in certain devices there is an increased probability of failure of the battery test circuit components of the UPS-module. Therefore the battery circuit for the UPS has been modified to make the UPS more robust.
Recall action type:Correction
Recall action:Product to be exchanged
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:9/3/2017