Detail of Recall

Type of Product:Device
Medsafe Reference:21191
Brand Name:Phadia EliA Sample Diluent
Model:
Affected: EliA Sample Diluent
Software version:
Recalling Organisation:Abacus ALS Limited
55b Richard Pearse Drive
Airport Oaks
Auckland 2022
Contact Information:0800 222 170
Manufacturer:
Issue:There have been reports of instrument malfunctions for Phadia 250 due to deformed bottles. The deformation of the bottle can cause an erroneous volume detection by the instrument which will result in insufficient aspiration of sample diluent, thus generating insufficient dilution of patient samples.
Recall action type:Recall
Recall action:Product to be destroyed
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/3/2017