Detail of Recall

Type of Product:Device
Medsafe Reference:21190
Brand Name:Siemens Dimension Vista 500 Intelligent Lab System and Dimension Vista® 1500 Intelligent Lab System
Model:
Affected: Catalogue numbers 10488224 and 10444802
Software version:
Recalling Organisation:Siemens Healthcare (NZ) Ltd
Millennium Centre, Part Level 2, Building A
600 Great South Road
Ellerslie
AUCKLAND 1051
Contact Information:0800 310 300
Manufacturer:Siemens Healthcare Diagnostics Inc
Issue:The manufacturer has confirmed that in isolated cases when Enzymatic Creatinine (ECREA) is processed immediately after the weekly automated Acid Clean routine during Probe Test, there is the remote potential for an elevation of greater than 15 percent in the ECREA result.
Recall action type:Correction
Recall action:Manufacturer to issue advice regarding use
Level of recall:Hospital/Pharmacy/Laboratory
Date Commenced:8/3/2017