Detail of Recall
| Type of Product: | Device |
| Medsafe Reference: | 21190 |
| Brand Name: | Siemens Dimension Vista 500 Intelligent Lab System and Dimension Vista® 1500 Intelligent Lab System |
| Model: | |
| Affected: | Catalogue numbers 10488224 and 10444802 |
| Software version: | |
| Recalling Organisation: | Siemens Healthcare (NZ) Ltd Millennium Centre, Part Level 2, Building A 600 Great South Road Ellerslie AUCKLAND 1051 |
| Contact Information: | 0800 310 300 |
| Manufacturer: | Siemens Healthcare Diagnostics Inc |
| Issue: | The manufacturer has confirmed that in isolated cases when Enzymatic Creatinine (ECREA) is processed immediately after the weekly automated Acid Clean routine during Probe Test, there is the remote potential for an elevation of greater than 15 percent in the ECREA result. |
| Recall action type: | Correction |
| Recall action: | Manufacturer to issue advice regarding use |
| Level of recall: | Hospital/Pharmacy/Laboratory |
| Date Commenced: | 8/3/2017 |